Ad ML-50 User Manual Page 44

  • Download
  • Add to my manuals
  • Print
  • Page
    / 74
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 43
42 MS-70 / MX-50 / MF-50 / ML-50
8. Check Function
8.1. Self-Check Function
Use the self-check function to check whether there is an error or inaccurate result.
During the check, the heater is turned on and the temperature sensor is checked.
Caution
Do not put flammable matter near the analyzer.
Do not put anything on the heater cover.
8.1.1. Operation
1. Display the gram unit (of the weighing mode).
2. Press and hold the PROGRAM key to
display CH .
3. Put the breeze break ring, pan support, pan
handle and sample pan in order. (Do not put
a sample on the pan.)
Close the heater cover.
Press the ENTER to start the check.
If Clo5e is displayed, the heater cover is
not closed. When it is closed, the self-check
function is started.
4. The check function needs approximately one
minute.
Good result... Displays
CH pa55 , sounds
buzzer and returns to weighing
mode automatically.
Error ............. The buzzer sounds and an error
code is displayed.
Refer to 14.5. Error Message for details.
Example:
CH no
error 0
Ht err
0.000g
CH
CH . . . .
0.000g
Preparation
CH pa55
Checking
Do not put sample
Press and hold
Good result
Weighing mode
Page view 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 73 74

Comments to this Manuals

No comments